Human Mesothelin ELISA Kit(one step)

Key features and details

  • Specification: 96 Test
  • Sensitivity: 0.16 ng/ml (50 μl);0.23 ng/ml (10 μl);
  • Standard Curve Range: 1.37~1000 ng/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 60min
  • Brand:
CAT.NO. : AEHY0152
US$ Please choose
US$ Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
Product Details
Background
Mesothelin, also known as CAK1 and ERC, is derived from a precursor that also includes Megakaryocyte Potentiating Factor (MPF). Following cleavage of the precursor, Mesothelin remains attached to the cell surface via a GPI linkage. Mesothelin is normally expressed on mesothelial cells in the pleura, pericardium, and peritoneum as well as in the developing and postnatal pancreas. It is upregulated in mesotheliomas and a range of carcinomas and adenomas. Mesothelin promotes tumor cell proliferation, migration, anchorage-independent growth, and tumor progression.
Typical data

ng/ml

O.D.

Average

Corrected

0.00

0.0128

0.0150

0.0139

1.37

0.0231

0.0235

0.0233

0.0094

4.12

0.0455

0.0442

0.0449

0.0310

12.35

0.1127

0.1187

0.1157

0.1018

37.04

0.3142

0.3075

0.3109

0.2970

111.11

0.8870

0.9468

0.9169

0.9030

333.33

2.5550

2.4420

2.4985

2.4846

1000.00

4.3296

4.4273

4.3785

4.3646

Precision

Intra-assay Precision

Inter-assay Precision

Sample Number

S1

S2

S3

S1

S2

S3

22

22

22

6

6

6

Average(ng/ml)

22.2

107.5

332.8

19.3

97.4

295.0

Standard Deviation

0.8

4.5

13.4

0.8

3.5

9.2

Coefficient of Variation(%)

3.5

4.2

4.0

3.9

3.5

3.1

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human Mesothelin into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 85% to 103% with an overall mean recovery of 94%.
Sample Values
Sample Matrix Sample Evaluated Range (ng/ml) Detectable (%) Mean of Detectable (ng/ml)
Serum 30 51.87-224.14 100 119.25

Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of Mesothelin in this assay. No medical histories were available for the donors.

New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.