E.coli Residual Total RNA Detection Kit (RT-PCR)

Key features and details

  • Detection Range: 2 - 2×10⁴ fg/μL
  • Limit of Quantitation: 2.0 fg/μL
  • Limit of Detection (LOD): 0.5 fg/μL
  • Precision: CV% ≤ 15%
  • Brand:
CAT.NO. : ARK0222
US$ Please choose
US$ Please choose
Trail, Bulk size or Custom requests Please contact us
Product Details
Background
This kit is designed for the quantitative detection of residual E.coli total RNA in various biological products to improve control quality of nucleic acid.
This kit adopts the principle of the RT-PCR fluorescent probe, combining reverse transcription PCR technology and fluorescent probe method, to realize one-step quantitative detection.
Overview
Parameter Specification
Assay Range 2 - 2×10⁴ fg/μL
Limit of Quantitation 2.0 fg/μL
Limit of Detection 0.5 fg/μL
Precision CV% ≤ 15%
Standard Curve
Performance
Concentration
(fg/uL)
Log10
Concentration
Ct Value(1) Ct Value(2) Ct Value(3) Ct Mean Value Recovery rate
2.00E+04 4.30 15.30 15.23 15.21 15.24 0.30%
2.00E+03 3.30 18.39 18.28 18.18 18.28 0.56%
2.00E+02 2.30 21.86 21.82 21.78 21.82 0.18%
2.00E+01 1.30 25.22 25.22 25.09 25.18 0.31%
2.00E+00 0.30 28.55 28.58 28.43 28.52 0.29%
Amplification efficiency 99.1%
Content
Component Vials Volume per Vial Storage Condition
One Step RT-qPCR Buffer 1 1 mL -20°C
One Step Enzyme Mix 1 100 µL -20°C
E. coli RNA Primer & Probe Mix 1 370 µL -20°C
E. coli RNA Quantitative Standard 1 25 µL -20°C
RNA IPC Primer & Probe Mix 1 370 µL -20°C
ROX High 1 50 µL -20°C
ROX Low 1 50 µL -20°C
RNA Diluent 2 1 mL Room Temperature
Storage
For long term storage, store at -20°C, avoiding freeze/thaw cycles.
Note
For Research Use Only. Not for diagnostic, therapeutics, prophylactic or in vivo use.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.