Human Nectin-4 ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 0.15 pg/ml (50 μl);5.23 pg/ml (10 μl)
  • Standard Curve Range: 2.74~2000 pg/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AEH0151
US$ Please choose
US$ Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
Product Details
Background
Nectin-4 is a type I transmembrane glycoprotein belonging to the Nectin family of Ig superfamily proteins. It is both a homophilic and heterophilic (with Nectin-1) cell adhesion molecule that is expressed in the embryo and in breast carcinoma. A soluble form of Nectin-4 is generated from the membrane protein via the action of TACE/ADAM-17. The extracellular domain of human Nectin-4 shares 90% and 92% amino acid sequence homology with the corresponding regions of mouse and rat Nectin-4, respectively.
Typical data

pg/ml

O.D.

Average

Corrected

0.00

0.0126

0.0126

0.0126

2.74

0.0216

0.0205

0.0211

0.0085

8.23

0.0338

0.0367

0.0353

0.0227

24.69

0.0781

0.0753

0.0767

0.0641

74.07

0.2027

0.2158

0.2093

0.1967

222.22

0.6049

0.6287

0.6168

0.6042

666.67

1.7850

1.8030

1.7940

1.7814

2000.00

3.6480

3.6120

3.6300

3.6174

Precision

Intra-assay Precision

Inter-assay Precision

Sample Number

S1

S2

S3

S1

S2

S3

22

22

22

6

6

6

Average(pg/ml)

49.8

240.7

861.5

44.4

211.8

614.5

Standard Deviation

1.9

11.5

27.1

1.3

12.3

35.4

Coefficient of Variation(%)

3.8

4.8

3.1

3.0

5.8

5.8

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human Nectin-4 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 100% to 117% with an overall mean recovery of 109%.
Sample Values
Sample Matrix Sample Evaluated Range (pg/ml) Detectable (%) Mean of Detectable (pg/ml)
Serum 30 56.23-451.61 100 226.49
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of Nectin-4 in this assay. No medical histories were available for the donors.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.