Human CD63 ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 0.24 pg/ml (50 μl);6.71 pg/ml (10 μl);
  • Standard Curve Range: 6.86~5000 pg/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AEH0167
US$ Please choose
US$ Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
Product Details
Background
CD63, also known as LAMP-3 or ME491 (melanoma-associated antigen), is a 30-60 kDa member of the tetraspanin superfamily of protein trafficking proteins. CD63 is ubiquitously expressed and found in late endocytic vesicles, but following cell activation is also present on the plasma membrane. Interaction of CD63 with other membrane proteins or adaptors regulates cell activities such as adhesion, migration and degranulation. Extracellular regions of human CD63 share 67% and 65% amino acid sequence identity with mouse and rat CD63, respectively.
Typical data

pg/ml

O.D.

Average

Corrected

0.00

0.0151

0.0128

0.0140

6.86

0.0286

0.0280

0.0283

0.0144

20.58

0.0572

0.0582

0.0577

0.0438

61.73

0.1475

0.1458

0.1467

0.1327

185.19

0.4194

0.3984

0.4089

0.3950

555.56

1.2120

1.1120

1.1620

1.1481

1666.67

2.6950

2.5740

2.6345

2.6206

5000.00

3.8760

3.8910

3.8835

3.8696

Precision

Intra-assay Precision

Inter-assay Precision

Sample Number

S1

S2

S3

S1

S2

S3

22

22

22

6

6

6

Average(pg/ml)

336.6

1807.6

5114.7

325.3

1844.1

5419.5

Standard Deviation

10.2

76.5

268.5

11.8

98.5

124.0

Coefficient of Variation(%)

3.0

4.2

5.2

3.6

5.3

2.3

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human CD63 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 93% to 119% with an overall mean recovery of 105%.
Sample Values
Sample Matrix Sample Evaluated Range (pg/ml) Detectable (%) Mean of Detectable (pg/ml)
Serum 30 570.74-1262.25 100 787.53
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of CD63 in this assay. No medical histories were available for the donors.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.