Human VEGF ELISA Kit
| pg/ml | O.D. | Average | Corrected | |
| 0.00 | 0.0275 | 0.0264 | 0.0270 | |
| 31.25 | 0.1190 | 0.1082 | 0.1136 | 0.0867 |
| 62.50 | 0.1991 | 0.2040 | 0.2016 | 0.1746 |
| 125.00 | 0.3159 | 0.2863 | 0.3011 | 0.2742 |
| 250.00 | 0.5521 | 0.6045 | 0.5783 | 0.5514 |
| 500.00 | 1.1560 | 1.2260 | 1.1910 | 1.1641 |
| 1000.00 | 2.3170 | 2.2130 | 2.2650 | 2.2381 |
| 2000.00 | 3.6280 | 3.4990 | 3.5635 | 3.5366 |
|
Intra-assay Precision |
Inter-assay Precision |
|||||
|
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
|
22 |
22 |
22 |
6 |
6 |
6 |
|
|
Average(pg/ml) |
53.2 |
229.1 |
694.5 |
31.6 |
160.6 |
438.1 |
|
Standard deviation |
3.6 |
13.3 |
38.4 |
2.1 |
8.2 |
13.6 |
|
Coefficient of variation(%) |
6.8 |
5.8 |
5.5 |
6.5 |
5.1 |
3.1 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human VEGF/VEGFA into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 94% to 119% with an overall mean recovery of 105%.
|
Sample Matrix |
Sample Evaluated |
Range (pg/ml) |
Detectable % |
Mean of Detectable (pg/ml) |
|
Serum |
30 |
25.21-450.32 |
100.0 |
185.52 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of VEGF in this assay. No medical histories were available for the donors.