Human Neuropilin-1/NRP1 ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 0.02 ng/ml (50 μl);0.03 ng/ml (10 μl);
  • Standard Curve Range: 0.27~200 ng/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 60min
  • Brand:
CAT.NO. : AEHY0175
US$ Please choose
US$ Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
Product Details
Background
Neuropilin-1 is a type I transmembrane protein that is expressed in the developing nervous system and by endothelial and tumor cells. Neuropilin-1 binds members of the class III secreted semaphorin subfamily as well as some isoforms of VEGF family proteins. The amino acid sequence of rat Neuropilin-1 extracellular domain is 98% and 93% identical to that of mouse and human Neuropilin-1, respectively.
Typical data

ng/ml

O.D.

Average

Corrected

0.00

0.0064

0.0061

0.0063

0.27

0.0235

0.0214

0.0225

0.0162

0.82

0.0624

0.0543

0.0584

0.0521

2.47

0.1677

0.1621

0.1649

0.1587

7.41

0.4859

0.4380

0.4620

0.4557

22.22

1.4320

1.3170

1.3745

1.3683

66.67

3.5090

3.4470

3.4780

3.4718

200.00

4.4662

4.5095

4.4879

4.4816

Precision

Intra-assay Precision

Inter-assay Precision

Sample Number

S1

S2

S3

S1

S2

S3

22

22

22

6

6

6

Average(ng/ml)

7.6

38.8

148.5

6.5

33.6

102.7

Standard Deviation

0.3

1.8

2.7

0.4

2.0

2.5

Coefficient of Variation(%)

3.6

4.5

1.8

6.0

6.0

2.5

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human Neuropilin-1/NRP1 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 80% to 125% with an overall mean recovery of 97%.
Sample Values
Sample Matrix Sample Evaluated Range (ng/ml) Detectable (%) Mean of Detectable (ng/ml)
Serum 30 407.50-604.49 100 514.05

Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of Neuropilin-1/NRP1 in this assay. No medical histories were available for the donors.

New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.
© AREX 2024. All Rights Reserved. Sitemap
0.256857s