Human CD62E ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 1.02 pg/ml (50 μl);5.13 pg/ml (10 μl);
  • Standard Curve Range: 8.23~6000 pg/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AEH0198
US$ Please choose
US$ Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
Product Details
Background
E-Selectin, also known as Endothelial Leukocyte Adhesion Molecule 1 (ELAM-1), and Leukocyte-Endothelial Cell Adhesion Molecule 2 (LECAM-2), has been designated CD62E. It is a cell-surface glycoprotein expressed by activated endothelium and it mediates the adhesion of blood neutrophils.
Typical data

pg/ml

O.D.

Average

Corrected

0.00

0.0179

0.0184

0.0182

8.23

0.0352

0.0343

0.0348

0.0166

24.69

0.0677

0.0684

0.0681

0.0499

74.07

0.1695

0.1696

0.1696

0.1514

222.22

0.4737

0.4691

0.4714

0.4533

666.67

1.2900

1.3210

1.3055

1.2874

2000.00

3.1610

3.2220

3.1915

3.1734

6000.00

4.4674

4.4727

4.4701

4.4519

Precision

Intra-assay Precision

Inter-assay Precision

Sample Number

S1

S2

S3

S1

S2

S3

22

22

22

6

6

6

Average(pg/ml)

163.0

726.7

2062.0

124.4

594.9

1727.9

Standard Deviation

9.2

29.1

128.8

5.2

31.9

74.36

Coefficient of Variation(%)

5.6

4.0

6.2

4.2

5.4

4.3

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human CD62E into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 85% to 115% with an overall mean recovery of 105%.
Sample Values
Sample Matrix Sample Evaluated Range (pg/ml) Detectable (%) Mean of Detectable (pg/ml)
Serum 30 8.16-43.12 100 19.69

Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of CD62E in this assay. No medical histories were available for the donors.

New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.