Rat IL-4 ELISA Kit
Sensitivity | 17.29 pg/ml (50 μl) |
Standard Curve Range | 2.74-2000 pg/ml |
Standard Curve Gradient | 7 Points/3 Folds |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | B-cell stimulatory factor 1, Binetrakin, Lymphocyte stimulatory factor 1, Pitrakinra |
Gene Symbol | Il4, Il4e12 |
Entrez Gene | 287287 (Rat) |
SwissProt | P20096 (Rat) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.105</td><td>0.104</td><td>0.1045</td><td></td></tr><tr><td>2.74</td><td>0.1086</td><td>0.1036</td><td>0.1061</td><td>0.0016</td></tr><tr><td>8.23</td><td>0.11</td><td>0.108</td><td>0.109</td><td>0.0045</td></tr><tr><td>24.69</td><td>0.1382</td><td>0.1433</td><td>0.1408</td><td>0.0767</td></tr><tr><td>74.07</td><td>0.2812</td><td>0.2953</td><td>0.2883</td><td>0.2242</td></tr><tr><td>222.22</td><td>0.6422</td><td>0.6543</td><td>0.6483</td><td>0.5842</td></tr><tr><td>666.67</td><td>1.5432</td><td>1.5153</td><td>1.5293</td><td>1.4652</td></tr><tr><td>2000</td><td>3.0881</td><td>3.1932</td><td>3.1407</td><td>3.0766</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 33.4 | 172.8 | 557.4 | 32.8 | 175.6 | 561.2 |
| Standard Deviation | 1.8 | 8.1 | 26.8 | 2.3 | 12.1 | 34.8 |
| Coefficient of Variation(%) | 5.4 | 4.7 | 4.8 | 7 | 6.9 | 6.2 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested Twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of rat IL-4 into health rat serum sample. The un-spiked serum was used as blank in these experiments. The recovery ranged from 93% to 115% with an overall mean recovery of 105%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | n.d. | 0 | 0 |
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.