Mouse/Rat VEGFR3/FLT4 ELISA Kit (one step)
Sensitivity | 0.18 ng/ml (10 μl) |
Standard Curve Range | 7.81-500 ng/ml |
Standard Curve Gradient | 7 Points |
Number of Incubations | 2 |
Detectable Sample | serum, plasma |
Sample Volume | 10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 60min |
Alternative Names | Fms-like tyrosine kinase 4, Tyrosine-protein kinase receptor FLT4 |
Gene Symbol | Flt4, Chy, Flt-4, VEGFR-3, VEGFR3 |
Entrez Gene | 14257 (Mouse) |
SwissProt | P35917 (Mouse) |

<table><tr><td>ng/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0064</td><td>0.0066</td><td>0.0066</td><td></td></tr><tr><td>7.81</td><td>0.0530</td><td>0.0511</td><td>0.0521</td><td>0.0455</td></tr><tr><td>15.63</td><td>0.1109</td><td>0.1045</td><td>0.1077</td><td>0.1011</td></tr><tr><td>31.25</td><td>0.2141</td><td>0.2223</td><td>0.2182</td><td>0.2116</td></tr><tr><td>62.50</td><td>0.4542</td><td>0.4671</td><td>0.4607</td><td>0.4541</td></tr><tr><td>125.00</td><td>0.9015</td><td>0.9226</td><td>0.9121</td><td>0.9055</td></tr><tr><td>250.00</td><td>1.8140</td><td>1.8360</td><td>1.8250</td><td>1.8184</td></tr><tr><td>500.00</td><td>3.1490</td><td>3.1270</td><td>3.1380</td><td>3.1314</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(ng/ml) | 10.0 | 47.4 | 152.3 | 10.4 | 48.9 | 160.6 |
| Standard Deviation | 0.5 | 1.7 | 5.2 | 0.5 | 1.6 | 7.3 |
| Coefficient of Variation(%) | 4.8 | 3.7 | 3.4 | 4.3 | 3.2 | 4.6 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of mouse/rat VEGFR3/FLT4 into health mouse/rat serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 90% to 112% with an overall mean recovery of 101%.
| Sample Matrix | Sample Evaluated | Range (ng/ml) | Detectable (%) | Mean of Detectable (ng/ml) |
|---|---|---|---|---|
| Mouse Serum | 30 | 94.52-190.88 | 100 | 159.39 |
| Rat Serum | 30 | 90.56-139.67 | 100 | 111.39 |
Serum/Plasma – Thirty samples from apparently healthy mice and thirty samples from apparently healthy rats were evaluated for the presence of mouse/rat VEGFR3/FLT4 in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.