Mouse/Rat NF-H/NEFH ELISA Kit

Key features and details

  • Target: NF-H/NEFH
  • Reactivity: Mouse,Rat
  • Range: 156.25-10000 pg/ml
  • Sensitivity: 20.96 pg/ml (10 μl)
  • Assay type Sandwich Elisa
  • Storage at 2 ~8°C
  • Brand:
CAT.NO. : AEM0090
US$ Please choose
US$ Please choose
Size:
Trail, Bulk size or Custom requests Please contact us
Product Details
Background
Neurofilaments are the 10nm or intermediate filament proteins found specifically in neurons, and are composed predominantly of three major proteins called Neurofilament Light (NF-L), Neurofilament Medium (NF-M) and Neurofilament Heavy (NF-H). Neurofilament Heavy is the heavy or high molecular weight microfilament subunit and runs on SDS-PAGE gels in the range 180-220kDa, with some variation in different species. Neurofilament Heavy antibodies can be used to identify neurons and their processes in tissue sections and in tissue culture. Additionally, antibodies can also be used to study microfilament accumulations seen in many neurological diseases, such as Lou Geri's disease or Alzheimer's disease.
Overview

Sensitivity

20.96 pg/ml (10 μl)

Standard Curve Range

156.25-10000 pg/ml

Standard Curve Gradient

7 Points

Number of Incubations

2

Detectable Sample

Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc.

Sample Volume

10 μl

Type

Ready-to-Use

Operation Duration

120min

Alternative Names

Gene Symbol

Nefh

Entrez Gene

18026 (Mouse)

SwissProt

P19246 (Mouse)

*AREX continuously optimizes our products. Webpage content may not reflect the latest updates. For inquiries, please contact info@arexbio.com or your local distributor.
Data

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0342</td><td>0.0310</td><td>0.0326</td><td></td></tr><tr><td>156.25</td><td>0.0732</td><td>0.0589</td><td>0.0661</td><td>0.0335</td></tr><tr><td>312.50</td><td>0.0972</td><td>0.1015</td><td>0.0994</td><td>0.0668</td></tr><tr><td>625.00</td><td>0.1670</td><td>0.1800</td><td>0.1735</td><td>0.1409</td></tr><tr><td>1250.00</td><td>0.3089</td><td>0.3034</td><td>0.3062</td><td>0.2736</td></tr><tr><td>2500.00</td><td>0.6334</td><td>0.6254</td><td>0.6294</td><td>0.5968</td></tr><tr><td>5000.00</td><td>1.3010</td><td>1.3190</td><td>1.3100</td><td>1.2774</td></tr><tr><td>10000.00</td><td>2.5660</td><td>2.4370</td><td>2.5015</td><td>2.4689</td></tr></table>

Precision
Intra-assay Precision Inter-assay Precision
Sample Number S1 S2 S3 S1 S2 S3
22 22 22 6 6 6
Average(pg/ml) 174.9 965.4 2965.3 202.5 1044.1 3109.0
Standard Deviation 13.0 56.6 130.9 14.5 28.6 56.5
Coefficient of Variation(%) 7.4 5.9 4.4 7.2 2.7 1.8

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess inter-assay precision.

Spike Recovery

The spike recovery was evaluated by spiking 3 levels of mouse/rat NF-H/NEFH into health mouse/rat serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 74% to 124% with an overall mean recovery of 90%.

Sample Values
Sample Matrix Sample Evaluated Range (pg/ml) Detectable (%) Mean of Detectable (pg/ml)
Mouse Serum30n.d.-577.2940133.27
Rat Serum30n.d.-76.52676.52

n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.

Storage
Store components at 4°C or -20°C. Keep microplate wells sealed in a dry bag with desiccants. Do not expose test reagents to heat, sun or strong light during storage and usage. Please refer to the product user manual for detail temperatures of the components.
Note
For Research Use Only. Not for diagnostic, therapeutics, prophylactic or in vivo use.
FAQs
What is the principle of these ELISA kits?
The kits are based on the sandwich or competitive ELISA principle, using specific antibodies and enzyme-conjugated detection for quantitative measurement of the target analyte.
What sample types can be used with the ELISA kits?
Suitable samples typically include serum, plasma, cell culture supernatant, tissue homogenates, and other biological fluids. Please check the specific Datasheet for recommended sample types.
How should the kits be stored and what is the shelf life?
Store components at 4°C or -20°C. Keep microplate wells sealed in a dry bag with desiccants. Do not expose test reagents to heat, sun or strong light during storage and usage. Please refer to the product user manual for detail temperatures of the components.
What is the precision and reproducibility of the kits?
The kits are optimized for good intra- and inter-assay precision when performed according to the protocol. Coefficient of variation (CV) values are generally low.
Do you provide technical support or troubleshooting guidance?
Yes. We provide detailed Datasheets, Certificates of Analysis (CoA), and technical support to help with assay performance and troubleshooting.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.
© AREX 2024. All Rights Reserved. Sitemap