Mouse/Rat IGF1 ELISA Kit (one step)
Sensitivity | 0.10 pg/ml (50 μl);0.37 pg/ml (10 μl); |
Standard Curve Range | 1.23-300 ng/ml |
Standard Curve Gradient | 7 Points |
Number of Incubations | 2 |
Detectable Sample | serum, plasma |
Sample Volume | 50 μl/10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 60min |
Alternative Names | Insulin-like growth factor I, Mechano growth factor, Somatomedin-C |
Gene Symbol | Igf1, C730016P09Rik, Igf-1, Igf-I |
Entrez Gene | 16000 (Mouse) |
SwissProt | P05017 (Mouse) |

<table><tr><td>ng/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0054</td><td>0.0060</td><td>0.0057</td><td></td></tr><tr><td>1.23</td><td>0.0110</td><td>0.0105</td><td>0.0108</td><td>0.0051</td></tr><tr><td>3.07</td><td>0.0193</td><td>0.0197</td><td>0.0195</td><td>0.0138</td></tr><tr><td>7.68</td><td>0.0436</td><td>0.0484</td><td>0.0460</td><td>0.0403</td></tr><tr><td>19.20</td><td>0.1256</td><td>0.1296</td><td>0.1276</td><td>0.1219</td></tr><tr><td>48.00</td><td>0.4230</td><td>0.4038</td><td>0.4134</td><td>0.4077</td></tr><tr><td>120.00</td><td>1.3870</td><td>1.3390</td><td>1.3630</td><td>1.3573</td></tr><tr><td>300.00</td><td>3.3950</td><td>3.3240</td><td>3.3595</td><td>3.3538</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(ng/ml) | 6.4 | 28.9 | 120.6 | 6.7 | 30.4 | 127.4 |
| Standard Deviation | 0.2 | 1.1 | 5.9 | 0.3 | 1.3 | 5.5 |
| Coefficient of Variation(%) | 3.7 | 3.7 | 4.9 | 3.8 | 4.2 | 4.3 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested Twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of mouse/rat IGF-I/IGF-1 into health mouse/rat serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 81% to 119% with an overall mean recovery of 99%.
| Sample Matrix | Sample Evaluated | Range (ng/ml) | Detectable (%) | Mean of Detectable (ng/ml) |
|---|---|---|---|---|
| Serum | 30 | 64.64-353.03 | 100 | 179.59 |