Mouse RANKL/TRANCE/TNFSF11 ELISA Kit
Sensitivity | 0.36 pg/ml (10 μl) |
Standard Curve Range | 2.74-2000 pg/ml |
Standard Curve Gradient | 7 Points/3 Folds |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | Osteoclast differentiation factor, Osteoprotegerin ligand, Receptor activator of nuclear factor kappa-B ligand, TNF-related activation-induced cytokine |
Gene Symbol | Tnfsf11, Ly109l, ODF, OPGL, RANKL, Trance |
Entrez Gene | 21943 (Mouse) |
SwissProt | O35235 (Mouse) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0138</td><td>0.0142</td><td>0.0140</td><td></td></tr><tr><td>2.74</td><td>0.0234</td><td>0.0247</td><td>0.0241</td><td>0.0101</td></tr><tr><td>8.23</td><td>0.0475</td><td>0.0485</td><td>0.0480</td><td>0.0340</td></tr><tr><td>24.69</td><td>0.1120</td><td>0.1176</td><td>0.1148</td><td>0.1008</td></tr><tr><td>74.07</td><td>0.3230</td><td>0.3205</td><td>0.3218</td><td>0.3078</td></tr><tr><td>222.22</td><td>0.9004</td><td>0.8719</td><td>0.8862</td><td>0.8722</td></tr><tr><td>666.67</td><td>2.2855</td><td>2.2587</td><td>2.2721</td><td>2.2581</td></tr><tr><td>2000.00</td><td>4.1051</td><td>3.9973</td><td>4.0512</td><td>4.0372</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 60.20 | 211.59 | 633.72 | 62.06 | 191.99 | 568.45 |
| Standard Deviation | 3.50 | 8.89 | 30.71 | 3.23 | 10.65 | 28.61 |
| Coefficient of Variation(%) | 5.8 | 4.2 | 4.8 | 5.2 | 5.5 | 5.0 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested Twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of mouse RANKL/TRANCE/TNFSF11 into health mouse serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 90% to 111% with an overall mean recovery of 100%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 61.89-141.84 | 100 | 107.38 |
FAQs
New Products
Instruction