Mouse LIF ELISA Kit
Sensitivity | 2.77 pg/ml (10 μl) |
Standard Curve Range | 6.86-5000 pg/ml |
Standard Curve Gradient | 7 Points/3 Folds |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | Differentiation-stimulating factor, Melanoma-derived LPL inhibitor |
Gene Symbol | Lif |
Entrez Gene | 16878 (Mouse) |
SwissProt | P09056 (Mouse) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0337</td><td>0.0309</td><td>0.0323</td><td></td></tr><tr><td>6.86</td><td>0.0458</td><td>0.0440</td><td>0.0449</td><td>0.0126</td></tr><tr><td>20.58</td><td>0.0698</td><td>0.0664</td><td>0.0681</td><td>0.0358</td></tr><tr><td>61.73</td><td>0.1404</td><td>0.1345</td><td>0.1375</td><td>0.1052</td></tr><tr><td>185.19</td><td>0.3507</td><td>0.3391</td><td>0.3449</td><td>0.3126</td></tr><tr><td>555.56</td><td>0.9113</td><td>0.9009</td><td>0.9061</td><td>0.8738</td></tr><tr><td>1666.67</td><td>2.1080</td><td>2.0780</td><td>2.0930</td><td>2.0607</td></tr><tr><td>5000.00</td><td>3.6630</td><td>3.7060</td><td>3.6845</td><td>3.6522</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 102.3 | 501.8 | 1489.4 | 100.7 | 491.3 | 1511.1 |
| Standard Deviation | 3.2 | 20.1 | 62.0 | 3.6 | 15.2 | 44.2 |
| Coefficient of Variation(%) | 3.1 | 4.0 | 4.2 | 3.5 | 3.1 | 2.9 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested Twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of mouse LIF into health mouse serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 94% to 105% with an overall mean recovery of 100%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 133.17-389.78 | 100 | 261.5 |
Serum/Plasma – Thirty samples from apparently healthy mice were evaluated for the presence of CRP in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.