Mouse CD200/OX2 ELISA Kit
Sensitivity | 0.13 pg/ml (10 μl) |
Standard Curve Range | 15.63-1000 pg/ml |
Standard Curve Gradient | 7 Points |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | |
Gene Symbol | Cd200, OX-2, MRC |
Entrez Gene | 17470 (Mouse) |
SwissProt | O54901 (Mouse) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0113</td><td>0.0106</td><td>0.0110</td><td></td></tr><tr><td>15.63</td><td>0.0972</td><td>0.0980</td><td>0.0976</td><td>0.0867</td></tr><tr><td>31.25</td><td>0.1881</td><td>0.1881</td><td>0.1881</td><td>0.1772</td></tr><tr><td>62.50</td><td>0.3766</td><td>0.3581</td><td>0.3674</td><td>0.3564</td></tr><tr><td>125.00</td><td>0.7547</td><td>0.7285</td><td>0.7416</td><td>0.7307</td></tr><tr><td>250.00</td><td>1.4150</td><td>1.4170</td><td>1.4160</td><td>1.4051</td></tr><tr><td>500.00</td><td>2.5000</td><td>2.5770</td><td>2.5385</td><td>2.5276</td></tr><tr><td>1000.00</td><td>3.7960</td><td>3.7070</td><td>3.7515</td><td>3.7406</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 21.0 | 98.7 | 286.6 | 22.4 | 103.0 | 286.6 |
| Standard Deviation | 1.0 | 4.1 | 17.1 | 0.8 | 4.7 | 8.1 |
| Coefficient of Variation(%) | 4.6 | 4.1 | 6.0 | 3.8 | 4.6 | 2.8 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess inter-assay precision.
The spike recovery was evaluated by spiking 3 levels of mouse CD200/OX2 into health mouse serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 85% to 117% with an overall mean recovery of 105%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 2990.65-5523.47 | 100 | 4314.28 |
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.