Mouse CCL3/MIP-1 alpha ELISA Kit
Sensitivity | 0.51 pg/ml (10 μl) |
Standard Curve Range | 1.37-1000 pg/ml |
Standard Curve Gradient | 7 Points/3 Folds |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | G0/G1 switch regulatory protein 19-1, Macrophage inflammatory protein 1-alpha, PAT 464.1, SIS-beta, Small-inducible cytokine A3, Tonsillar lymphocyte LD78 alpha protein |
Gene Symbol | Ccl3, G0S19-1, LD78alpha, MIP-1alpha, MIP1-(a), MIP1-alpha |
Entrez Gene | 20302 (Mouse) |
SwissProt | P10855 (Mouse) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0159</td><td>0.0141</td><td>0.0150</td><td></td></tr><tr><td>1.37</td><td>0.0232</td><td>0.0210</td><td>0.0221</td><td>0.0071</td></tr><tr><td>4.12</td><td>0.0341</td><td>0.0302</td><td>0.0322</td><td>0.0172</td></tr><tr><td>12.35</td><td>0.0651</td><td>0.0658</td><td>0.0655</td><td>0.0505</td></tr><tr><td>37.04</td><td>0.1703</td><td>0.1707</td><td>0.1705</td><td>0.1555</td></tr><tr><td>111.11</td><td>0.4507</td><td>0.4318</td><td>0.4413</td><td>0.4263</td></tr><tr><td>333.33</td><td>1.2730</td><td>1.2200</td><td>1.2465</td><td>1.2315</td></tr><tr><td>1000.00</td><td>3.1870</td><td>3.2800</td><td>3.2335</td><td>3.2185</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 18.8 | 88.6 | 247.5 | 18.5 | 96.2 | 282.2 |
| Standard Deviation | 1.4 | 3.9 | 13.2 | 1.2 | 6.7 | 10.1 |
| Coefficient of Variation(%) | 7.3 | 4.4 | 5.3 | 6.6 | 6.9 | 3.6 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested Twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of mouse CCL3 into health mouse serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 106% to 139% with an overall mean recovery of 128%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 1.2-20.8 | 100 | 4.19 |
Serum/Plasma – Thirty samples from apparently healthy mice were evaluated for the presence of CCL5 in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
FAQs
New Products
Instruction