Mammalian CD62P/P-Selectin ELISA Kit
Sensitivity | 8.19 pg/ml (50 μl);70.99 pg/ml (10 μl); |
Standard Curve Range | 156.25-10000 pg/ml |
Standard Curve Gradient | 7 Points |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl/10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | CD62 antigen-like family member P, Granule membrane protein 140, Leukocyte-endothelial cell adhesion molecule 3, Platelet activation dependent granule-external membrane protein |
Gene Symbol | SELP, GMRP, GRMP |
Entrez Gene | 6403 (Human) |
SwissProt | P16109 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0444</td><td>0.0441</td><td>0.0443</td><td></td></tr><tr><td>156.25</td><td>0.1587</td><td>0.1492</td><td>0.1540</td><td>0.1097</td></tr><tr><td>312.50</td><td>0.2757</td><td>0.2684</td><td>0.2721</td><td>0.2278</td></tr><tr><td>625.00</td><td>0.4935</td><td>0.4962</td><td>0.4949</td><td>0.4506</td></tr><tr><td>1250.00</td><td>0.8932</td><td>0.8620</td><td>0.8776</td><td>0.8334</td></tr><tr><td>2500.00</td><td>1.4950</td><td>1.5650</td><td>1.5300</td><td>1.4858</td></tr><tr><td>5000.00</td><td>2.3990</td><td>2.4770</td><td>2.4380</td><td>2.3938</td></tr><tr><td>10000.00</td><td>3.3780</td><td>3.4450</td><td>3.4115</td><td>3.3673</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 199.2 | 1022.8 | 3359.7 | 161.6 | 695.0 | 2182.5 |
| Standard Deviation | 10.0 | 44.6 | 381.0 | 7.6 | 29.3 | 104.0 |
| Coefficient of Variation(%) | 5.0 | 4.4 | 11.3 | 4.7 | 4.2 | 4.8 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of mammalian CD62P/P-Selectin into health mammalian serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 80% to 115% with an overall mean recovery of 94%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Human Serum | 30 | 7.85-29.14 | 100 | 18.17 |
| Mouse Serum | 30 | 355.11-903.67 | 100 | 572.82 |
| Rat Serum | 30 | 26.18-72.12 | 100 | 46.71 |
| Monkey Serum | 1 | 4.32 | 100 | 4.32 |
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
FAQs
New Products
Instruction