Mammalian CD62P/P-Selectin ELISA Kit

Key features and details

  • Target: CD62P/P-Selectin
  • Reactivity: Mammalian
  • Range: 156.25-10000 pg/ml
  • Sensitivity: 8.19 pg/ml (50 μl);70.99 pg/ml (10 μl);
  • Assay type Sandwich Elisa
  • Storage at 2 ~8°C
  • Brand:
CAT.NO. : AEH0289
US$ Please choose
US$ Please choose
Size:
Trail, Bulk size or Custom requests Please contact us
Product Details
Background
E-Selectin and P-Selectin are related (44% amino acid sequence identity) adhesion molecules containing multiple short complement-like repeats and an N-terminal C‑type lectin domain in their extracellular regions. E-Selectin is expressed on activated vascular endothelial cells and P-Selectin occurs on activated platelets and endothelial cells.
Overview

Sensitivity

8.19 pg/ml (50 μl);70.99 pg/ml (10 μl);

Standard Curve Range

156.25-10000 pg/ml

Standard Curve Gradient

7 Points

Number of Incubations

2

Detectable Sample

Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc.

Sample Volume

50 μl/10 μl

Type

Fully Ready-to-Use

Operation Duration

120min

Alternative Names

CD62 antigen-like family member P, Granule membrane protein 140, Leukocyte-endothelial cell adhesion molecule 3, Platelet activation dependent granule-external membrane protein

Gene Symbol

SELP, GMRP, GRMP

Entrez Gene

6403 (Human)

SwissProt

P16109 (Human)

*AREX continuously optimizes our products. Webpage content may not reflect the latest updates. For inquiries, please contact info@arexbio.com or your local distributor.
Data

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0444</td><td>0.0441</td><td>0.0443</td><td></td></tr><tr><td>156.25</td><td>0.1587</td><td>0.1492</td><td>0.1540</td><td>0.1097</td></tr><tr><td>312.50</td><td>0.2757</td><td>0.2684</td><td>0.2721</td><td>0.2278</td></tr><tr><td>625.00</td><td>0.4935</td><td>0.4962</td><td>0.4949</td><td>0.4506</td></tr><tr><td>1250.00</td><td>0.8932</td><td>0.8620</td><td>0.8776</td><td>0.8334</td></tr><tr><td>2500.00</td><td>1.4950</td><td>1.5650</td><td>1.5300</td><td>1.4858</td></tr><tr><td>5000.00</td><td>2.3990</td><td>2.4770</td><td>2.4380</td><td>2.3938</td></tr><tr><td>10000.00</td><td>3.3780</td><td>3.4450</td><td>3.4115</td><td>3.3673</td></tr></table>

Precision
Intra-assay Precision Inter-assay Precision
Sample Number S1 S2 S3 S1 S2 S3
22 22 22 6 6 6
Average(pg/ml) 199.2 1022.8 3359.7 161.6 695.0 2182.5
Standard Deviation 10.0 44.6 381.0 7.6 29.3 104.0
Coefficient of Variation(%) 5.0 4.4 11.3 4.7 4.2 4.8

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery

The spike recovery was evaluated by spiking 3 levels of mammalian CD62P/P-Selectin into health mammalian serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 80% to 115% with an overall mean recovery of 94%.

Sample Values
Sample Matrix Sample Evaluated Range (pg/ml) Detectable (%) Mean of Detectable (pg/ml)
Human Serum307.85-29.1410018.17
Mouse Serum30355.11-903.67100572.82
Rat Serum3026.18-72.1210046.71
Monkey Serum14.321004.32

n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.

Storage
Store components at 4°C or -20°C. Keep microplate wells sealed in a dry bag with desiccants. Do not expose test reagents to heat, sun or strong light during storage and usage. Please refer to the product user manual for detail temperatures of the components.
Note
For Research Use Only. Not for diagnostic, therapeutics, prophylactic or in vivo use.
FAQs
What is the principle of these ELISA kits?
The kits are based on the sandwich or competitive ELISA principle, using specific antibodies and enzyme-conjugated detection for quantitative measurement of the target analyte.
What sample types can be used with the ELISA kits?
Suitable samples typically include serum, plasma, cell culture supernatant, tissue homogenates, and other biological fluids. Please check the specific Datasheet for recommended sample types.
How should the kits be stored and what is the shelf life?
Store components at 4°C or -20°C. Keep microplate wells sealed in a dry bag with desiccants. Do not expose test reagents to heat, sun or strong light during storage and usage. Please refer to the product user manual for detail temperatures of the components.
What is the precision and reproducibility of the kits?
The kits are optimized for good intra- and inter-assay precision when performed according to the protocol. Coefficient of variation (CV) values are generally low.
Do you provide technical support or troubleshooting guidance?
Yes. We provide detailed Datasheets, Certificates of Analysis (CoA), and technical support to help with assay performance and troubleshooting.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.
© AREX 2024. All Rights Reserved. Sitemap