Human von Willebrand Factor ELISA Kit (one step)
Sensitivity | 0.02 ng/ml (50 μl);0.69 ng/ml (10 μl); |
Standard Curve Range | 4.69-300 ng/ml |
Standard Curve Gradient | 7 Points |
Number of Incubations | 2 |
Detectable Sample | serum, plasma |
Sample Volume | 50 μl/10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 60min |
Alternative Names | F8VWF, VWD, von Willebrand factor, Factor VIII related antigen, coagulation factor VIII VWF |
Gene Symbol | VWF, F8VWF |
Entrez Gene | 7450 (Human) |
SwissProt | P04275 (Human) |

<table><tr><td>ng/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0064</td><td>0.0058</td><td>0.0061</td><td></td></tr><tr><td>4.69</td><td>0.0354</td><td>0.0345</td><td>0.0350</td><td>0.0293</td></tr><tr><td>9.38</td><td>0.0859</td><td>0.0789</td><td>0.0824</td><td>0.0798</td></tr><tr><td>18.75</td><td>0.2116</td><td>0.1990</td><td>0.2053</td><td>0.2055</td></tr><tr><td>37.50</td><td>0.5367</td><td>0.5317</td><td>0.5342</td><td>0.5306</td></tr><tr><td>75.00</td><td>1.3890</td><td>1.3180</td><td>1.3535</td><td>1.3829</td></tr><tr><td>150.00</td><td>2.5750</td><td>2.7090</td><td>2.6420</td><td>2.5689</td></tr><tr><td>300.00</td><td>3.6260</td><td>3.6480</td><td>3.6370</td><td>3.6199</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(ng/ml) | 5.7 | 32.8 | 151.8 | 5.3 | 31.3 | 113.6 |
| Standard Deviation | 0.2 | 1.7 | 10.4 | 0.1 | 2.0 | 7.7 |
| Coefficient of Variation(%) | 2.7 | 5.2 | 6.8 | 2.6 | 6.4 | 6.7 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested Twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human von Willebrand Factor into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 94% to 126% with an overall mean recovery of 109%.
| Sample Matrix | Sample Evaluated | Range (ng/ml) | Detectable (%) | Mean of Detectable (ng/ml) |
|---|---|---|---|---|
| Serum | 30 | 10.53-1936.87 | 100 | 1056.08 |
FAQs