Human SOST/Sclerostin ELISA Kit
Sensitivity | 0.01 pg/ml (50 μl);1.99 pg/ml (10 μl) |
Standard Curve Range | 8.19-2000 pg/ml |
Standard Curve Gradient | 7 Points |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl/10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | |
Gene Symbol | SOST |
Entrez Gene | 50964 (Human) |
SwissProt | Q9BQB4 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0122</td><td>0.0117</td><td>0.0120</td><td></td></tr><tr><td>8.19</td><td>0.0403</td><td>0.0400</td><td>0.0402</td><td>0.0282</td></tr><tr><td>20.48</td><td>0.0871</td><td>0.0849</td><td>0.0860</td><td>0.0741</td></tr><tr><td>51.20</td><td>0.2017</td><td>0.2058</td><td>0.2038</td><td>0.1918</td></tr><tr><td>128.00</td><td>0.5116</td><td>0.4851</td><td>0.4984</td><td>0.4864</td></tr><tr><td>320.00</td><td>1.1817</td><td>1.2118</td><td>1.1968</td><td>1.1848</td></tr><tr><td>800.00</td><td>2.6924</td><td>2.6519</td><td>2.6722</td><td>2.6602</td></tr><tr><td>2000.00</td><td>4.2647</td><td>4.3375</td><td>4.3011</td><td>4.2892</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 45.4 | 206.7 | 657.6 | 44.1 | 207.5 | 653.8 |
| Standard Deviation | 1.0 | 5.8 | 16.3 | 0.9 | 8.8 | 21.5 |
| Coefficient of Variation(%) | 2.2 | 2.8 | 2.5 | 2.1 | 4.2 | 3.3 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess inter-assay precision.
The spike recovery was evaluated by spiking 3 levels of human SOST/Sclerostin into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 81% to 109% with an overall mean recovery of 99%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 26.50-325.57 | 100 | 112.33 |
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.