Human Serpin A12 ELISA Kit
Sensitivity | 2.48 pg/ml (50 μl);19.23 pg/ml (10 μl) |
Standard Curve Range | 31.25-2000 pg/ml |
Standard Curve Gradient | 7 Points |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl/10 μl |
Type | Ready-to-Use |
Operation Duration | 120min |
Alternative Names | |
Gene Symbol | SERPINA12 |
Entrez Gene | 145264 (Human) |
SwissProt | Q8IW75 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0779</td><td>0.0784</td><td>0.0782</td><td></td></tr><tr><td>31.25</td><td>0.1476</td><td>0.1496</td><td>0.1486</td><td>0.0705</td></tr><tr><td>62.50</td><td>0.2088</td><td>0.2223</td><td>0.2156</td><td>0.1374</td></tr><tr><td>125.00</td><td>0.3183</td><td>0.3407</td><td>0.3295</td><td>0.2514</td></tr><tr><td>250.00</td><td>0.5592</td><td>0.5849</td><td>0.5721</td><td>0.4939</td></tr><tr><td>500.00</td><td>0.9686</td><td>1.0220</td><td>0.9953</td><td>0.9172</td></tr><tr><td>1000.00</td><td>1.5950</td><td>1.6860</td><td>1.6405</td><td>1.5624</td></tr><tr><td>2000.00</td><td>2.6530</td><td>2.7610</td><td>2.7070</td><td>2.6289</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 26.4 | 202.7 | 639.4 | 31.6 | 191.2 | 557.5 |
| Standard Deviation | 2.0 | 14.5 | 46.7 | 2.2 | 7.9 | 28.1 |
| Coefficient of Variation(%) | 7.5 | 7.1 | 7.3 | 7.0 | 4.1 | 5.0 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess inter-assay precision.
The spike recovery was evaluated by spiking 3 levels of human Serpin A12 into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 70% to 95% with an overall mean recovery of 80%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 6.64-218.03 | 100 | 48.88 |
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.