Human SAA1 ELISA Kit
Sensitivity | 0.30ng/ml (50 μl);1.56 ng/ml (10 μl); |
Standard Curve Range | 1.56-100 ng/ml |
Standard Curve Gradient | 7 Points |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl/10 μl |
Type | Ready-to-Use |
Operation Duration | 120min |
Alternative Names | PIG4, SAA, TP53I4, serum amyloid A-1 protein, serum amyloid A protein, tumor protein p53 inducible protein 4 |
Gene Symbol | SAA1 |
Entrez Gene | 6288 (Human) |
SwissProt | P0DJI8 (Human) |

<table><tr><td>ng/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0226</td><td>0.0227</td><td>0.0227</td><td></td></tr><tr><td>1.56</td><td>0.0334</td><td>0.0370</td><td>0.0352</td><td>0.0126</td></tr><tr><td>3.13</td><td>0.0579</td><td>0.0619</td><td>0.0599</td><td>0.0373</td></tr><tr><td>6.25</td><td>0.1442</td><td>0.1511</td><td>0.1477</td><td>0.1250</td></tr><tr><td>12.50</td><td>0.3880</td><td>0.4018</td><td>0.3949</td><td>0.3723</td></tr><tr><td>25.00</td><td>1.2466</td><td>1.2545</td><td>1.2506</td><td>1.2279</td></tr><tr><td>50.00</td><td>3.1485</td><td>3.0604</td><td>3.1045</td><td>3.0818</td></tr><tr><td>100.00</td><td>4.6630</td><td>4.7992</td><td>4.7311</td><td>4.7085</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(ng/ml) | 2.4 | 12.0 | 38.0 | 2.5 | 11.4 | 35.1 |
| Standard Deviation | 0.1 | 0.4 | 1.3 | 0.1 | 0.5 | 1.3 |
| Coefficient of Variation(%) | 5.6 | 3.0 | 3.5 | 2.9 | 4.1 | 3.7 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested Twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human SAA1 into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 83% to 117% with an overall mean recovery of 100%.
| Sample Matrix | Sample Evaluated | Range (ng/ml) | Detectable (%) | Mean of Detectable (ng/ml) |
|---|---|---|---|---|
| Serum | 30 | 24.57-213.30 | 100 | 118.34 |
FAQs
New Products
Instruction