Human Prealbumin/Transthyretin ELISA Kit (one step)
Sensitivity | 0.02 ng/ml (10 μl) |
Standard Curve Range | 3.13-200 ng/ml |
Standard Curve Gradient | 7 Points |
Number of Incubations | 2 |
Detectable Sample | serum, plasma |
Sample Volume | 10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 60min |
Alternative Names | ATTR, Prealbumin, TBPA |
Gene Symbol | TTR, PALB |
Entrez Gene | 7276 (Human) |
SwissProt | P02766 (Human) |

<table><tr><td>ng/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0095</td><td>0.0098</td><td>0.0097</td><td></td></tr><tr><td>3.13</td><td>0.0895</td><td>0.0823</td><td>0.0859</td><td>0.0763</td></tr><tr><td>6.25</td><td>0.1597</td><td>0.1545</td><td>0.1571</td><td>0.1475</td></tr><tr><td>12.50</td><td>0.3028</td><td>0.3084</td><td>0.3056</td><td>0.2960</td></tr><tr><td>25.00</td><td>0.6030</td><td>0.5964</td><td>0.5997</td><td>0.5901</td></tr><tr><td>50.00</td><td>1.1280</td><td>1.1260</td><td>1.1270</td><td>1.1174</td></tr><tr><td>100.00</td><td>1.9260</td><td>1.8610</td><td>1.8935</td><td>1.8839</td></tr><tr><td>200.00</td><td>3.0040</td><td>2.9850</td><td>2.9945</td><td>2.9849</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(ng/ml) | 4.2 | 19.3 | 58.0 | 4.5 | 19.9 | 59.2 |
| Standard Deviation | 0.2 | 0.9 | 2.8 | 0.1 | 0.6 | 2.4 |
| Coefficient of Variation(%) | 4.7 | 4.5 | 4.8 | 2.4 | 3.1 | 4.0 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human Prealbumin/Transthyretin into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 83% to 115% with an overall mean recovery of 97%.
| Sample Matrix | Sample Evaluated | Range (μg/ml) | Detectable (%) | Mean of Detectable (μg/ml) |
|---|---|---|---|---|
| Serum | 30 | 227.44-506.95 | 100 | 378.79 |