Human/Mouse/Rat FGF1 ELISA Kit
Sensitivity | 0.21 pg/ml (50 μl);0.98 pg/ml (10 μl) |
Standard Curve Range | 1.37-1000 pg/ml |
Standard Curve Gradient | 7 Points/3 Folds |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl/10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | |
Gene Symbol | FGF1, FGFA |
Entrez Gene | 2246 (Human) |
SwissProt | P05230 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0171</td><td>0.0159</td><td>0.0165</td><td></td></tr><tr><td>1.37</td><td>0.0238</td><td>0.0247</td><td>0.0243</td><td>0.0078</td></tr><tr><td>4.12</td><td>0.0405</td><td>0.0407</td><td>0.0406</td><td>0.0241</td></tr><tr><td>12.35</td><td>0.0853</td><td>0.0860</td><td>0.0857</td><td>0.0692</td></tr><tr><td>37.04</td><td>0.2135</td><td>0.2223</td><td>0.2179</td><td>0.2014</td></tr><tr><td>111.11</td><td>0.6743</td><td>0.6299</td><td>0.6521</td><td>0.6356</td></tr><tr><td>333.33</td><td>1.7250</td><td>1.5970</td><td>1.6610</td><td>1.6445</td></tr><tr><td>1000.00</td><td>3.6830</td><td>3.6440</td><td>3.6635</td><td>3.6470</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 18.8 | 96.3 | 320.4 | 22.3 | 117.8 | 344.7 |
| Standard Deviation | 0.6 | 5.4 | 16.5 | 0.8 | 4.1 | 11.7 |
| Coefficient of Variation(%) | 3.2 | 5.6 | 5.1 | 3.4 | 3.5 | 3.4 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess inter-assay precision.
The spike recovery was evaluated by spiking 3 levels of human FGF1 into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 92% to 112% with an overall mean recovery of 102%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Human Serum | 30 | n.d.-5.34 | 6.25 | 0.33 |
| Rat Serum | 30 | n.d.-138.45 | 50 | 15.90 |
| Mouse Serum | 30 | n.d.-378.55 | 31.25 | 32.65 |
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
FAQs
New Products
Instruction