Human/Monkey VEGFR3/FLT4 ELISA Kit (one step)
Sensitivity | 0.10 ng/ml (10 μl) |
Standard Curve Range | 7.81-500 ng/ml |
Standard Curve Gradient | 7 Points |
Number of Incubations | 1 |
Detectable Sample | serum, plasma |
Sample Volume | 10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 60min |
Alternative Names | Fms-like tyrosine kinase 4, Tyrosine-protein kinase receptor FLT4 |
Gene Symbol | FLT4, VEGFR3 |
Entrez Gene | 2324 (Human) |
SwissProt | P35916 (Human) |

<table><tr><td>ng/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0035</td><td>0.0032</td><td>0.0034</td><td></td></tr><tr><td>7.81</td><td>0.0605</td><td>0.0614</td><td>0.0610</td><td>0.0578</td></tr><tr><td>15.63</td><td>0.1381</td><td>0.1410</td><td>0.1396</td><td>0.1364</td></tr><tr><td>31.25</td><td>0.2995</td><td>0.2867</td><td>0.2931</td><td>0.2899</td></tr><tr><td>62.50</td><td>0.5641</td><td>0.6069</td><td>0.5855</td><td>0.5823</td></tr><tr><td>125.00</td><td>1.0540</td><td>1.0820</td><td>1.0680</td><td>1.0648</td></tr><tr><td>250.00</td><td>1.7850</td><td>1.6480</td><td>1.7165</td><td>1.7133</td></tr><tr><td>500.00</td><td>2.4030</td><td>2.4070</td><td>2.4050</td><td>2.4018</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(ng/ml) | 7.3 | 38.1 | 128.0 | 8.3 | 40.4 | 128.0 |
| Standard Deviation | 0.3 | 1.8 | 6.3 | 0.4 | 1.8 | 6.1 |
| Coefficient of Variation(%) | 4.6 | 4.7 | 4.9 | 4.9 | 4.5 | 4.8 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human/monkey VEGFR3/FLT4 into health human/monkey serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 91% to 117% with an overall mean recovery of 101%.
| Sample Matrix | Sample Evaluated | Range (ng/ml) | Detectable (%) | Mean of Detectable (ng/ml) |
|---|---|---|---|---|
| Human Serum | 30 | 26.62-155.63 | 100 | 101.03 |
| Monkey Serum | 2 | 100.08-251.00 | 100 | 175.54 |
Serum/Plasma – Thirty samples from apparently healthy volunteers and two samples from healthy monkeys were evaluated for the presence of human/monkey VEGFR3/FLT4 in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
FAQs
New Products