Human MMP-13 ELISA Kit
Sensitivity | 5.09 pg/ml (50 μl);10.80 pg/ml (10 μl) |
Standard Curve Range | 78.13-5000 pg/ml |
Standard Curve Gradient | 7 Points |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl/10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | Matrix metalloproteinase-13 |
Gene Symbol | MMP13 |
Entrez Gene | 4322 (Human) |
SwissProt | P45452 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0792</td><td>0.0843</td><td>0.0818</td><td></td></tr><tr><td>78.13</td><td>0.1480</td><td>0.1524</td><td>0.1502</td><td>0.0685</td></tr><tr><td>156.25</td><td>0.2363</td><td>0.2171</td><td>0.2267</td><td>0.1450</td></tr><tr><td>312.50</td><td>0.3585</td><td>0.3677</td><td>0.3631</td><td>0.2814</td></tr><tr><td>625.00</td><td>0.6069</td><td>0.6087</td><td>0.6078</td><td>0.5261</td></tr><tr><td>1250.00</td><td>1.1020</td><td>1.0940</td><td>1.0980</td><td>1.0163</td></tr><tr><td>2500.00</td><td>1.8900</td><td>1.9660</td><td>1.9280</td><td>1.8463</td></tr><tr><td>5000.00</td><td>3.3930</td><td>3.3560</td><td>3.3745</td><td>3.2928</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 56.9 | 372.0 | 1283.4 | 77.7 | 432.4 | 1499.5 |
| Standard Deviation | 3.3 | 22.3 | 74.8 | 2.8 | 13.2 | 38.0 |
| Coefficient of Variation(%) | 5.8 | 6.0 | 5.8 | 3.6 | 3.1 | 2.5 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested Twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human MMP-13 into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 92% to 117% with an overall mean recovery of 106%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 5.75-44.20 | 100 | 21.14 |
FAQs
New Products
Instruction