Human IL4 TR-FRET Detection Kit
Assay Time | 2 hours at room temperature |
Detection Range | 47.7 – 5,000 pg/mL |
Limit of Detection (LoD) | 7.1 |
Limit of Quantification (LoQ) | 47.7 |
Specificity | Human IL4 |
Sample Type | Cell culture supernatant, serum, plasma |
Storage | ≤ -60 °C / 2-8 °C |

<table><tr><td>Standard</td><td>Concentration (pg/mL)</td><td>TR-FRET Ratio</td><td>Delta Ratio</td><td>CV%</td></tr><tr><td>STD-7</td><td>5,000.0</td><td>19,105</td><td>18,300</td><td>0.5</td></tr><tr><td>STD-6</td><td>2,272.7</td><td>12,500</td><td>11,695</td><td>3.6</td></tr><tr><td>STD-5</td><td>1,033.1</td><td>7,089</td><td>6,284</td><td>3.0</td></tr><tr><td>STD-4</td><td>469.6</td><td>4,213</td><td>3,408</td><td>0.3</td></tr><tr><td>STD-3</td><td>213.4</td><td>2,638</td><td>1,833</td><td>0.3</td></tr><tr><td>STD-2</td><td>97.0</td><td>1,642</td><td>837</td><td>2.8</td></tr><tr><td>STD-1</td><td>44.1</td><td>1,153</td><td>347</td><td>3.7</td></tr><tr><td>STD-0</td><td>0.0</td><td>805</td><td>0</td><td>0.6</td></tr></table>
•Low sample consumption (16 µL per well).
•Wide quantifiable range and high reproducibility.
| Parameter | Diluent buffer | RPMI1640 + 10% FBS | DMEM + 10% FBS | 100% FBS |
| Assay range (pg/mL) | 47.7 – 5,000 pg/mL | |||
| LoD (Std 0 mean + 2×SD) | 7.1 pg/mL | 12.7 pg/mL | 8.5 pg/mL | 29.2 pg/mL |
| LoQ (Std 0 mean + 10×SD) | 47.7 pg/mL | 56.7 pg/mL | 51.5 pg/mL | 108.7 pg/mL |
| Incubation time | 2 hours at room temperature | |||