Human IL1RL1/DER4/ST2 ELISA Kit
Sensitivity | 1.59 pg/ml (50 μl);8.53 pg/ml (10 μl); |
Standard Curve Range | 13.72-10000 pg/ml |
Standard Curve Gradient | 7 Points/3 Folds |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 10 μl/50 μl |
Type | Ready-to-Use |
Operation Duration | 120min |
Alternative Names | Protein ST2 |
Gene Symbol | IL1RL1, DER4, ST2, T1 |
Entrez Gene | 9173 (Human) |
SwissProt | Q01638 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0084</td><td>0.0073</td><td>0.0079</td><td></td></tr><tr><td>13.72</td><td>0.0170</td><td>0.0168</td><td>0.0169</td><td>0.0091</td></tr><tr><td>41.15</td><td>0.0341</td><td>0.0368</td><td>0.0355</td><td>0.0276</td></tr><tr><td>123.46</td><td>0.0847</td><td>0.0891</td><td>0.0869</td><td>0.0791</td></tr><tr><td>370.37</td><td>0.2346</td><td>0.2441</td><td>0.2394</td><td>0.2315</td></tr><tr><td>1111.11</td><td>0.5879</td><td>0.5783</td><td>0.5831</td><td>0.5753</td></tr><tr><td>3333.33</td><td>1.5660</td><td>1.5200</td><td>1.5430</td><td>1.5352</td></tr><tr><td>10000.00</td><td>3.1700</td><td>3.1730</td><td>3.1715</td><td>3.1637</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 211.53 | 980.92 | 2967.31 | 245.35 | 1173.26 | 3093.95 |
| Standard Deviation | 9.43 | 46.15 | 168.39 | 17.87 | 72.26 | 172.49 |
| Coefficient of Variation(%) | 4.5 | 4.7 | 5.7 | 7.3 | 6.2 | 5.6 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested Twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human IL1RL1 into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 85% to 120% with an overall mean recovery of 108%.
| Sample Matrix | Sample Evaluated | Range (ng/ml) | Detectable (%) | Mean of Detectable (ng/ml) |
|---|---|---|---|---|
| Serum | 30 | 17.61-61.57 | 100.0 | 28.48 |
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.