Human IL-6 Super Sensitivity ELISA Kit
Sensitivity | 0.11 pg/ml (50 μl);0.33 pg/ml (10 μl) |
Standard Curve Range | 0.14-100 pg/ml |
Standard Curve Gradient | 7 Points/3 Folds |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl/10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | B-cell stimulatory factor 2, CTL differentiation factor, Hybridoma growth factor, Interferon beta-2 |
Gene Symbol | IL6, IFNB2 |
Entrez Gene | 3569 (Human) |
SwissProt | P05231 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0565</td><td>0.0573</td><td>0.0569</td><td></td></tr><tr><td>0.14</td><td>0.0621</td><td>0.0635</td><td>0.0628</td><td>0.0059</td></tr><tr><td>0.41</td><td>0.0701</td><td>0.0717</td><td>0.0709</td><td>0.014</td></tr><tr><td>1.23</td><td>0.0919</td><td>0.0878</td><td>0.0899</td><td>0.033</td></tr><tr><td>3.7</td><td>0.1406</td><td>0.1335</td><td>0.1371</td><td>0.0802</td></tr><tr><td>11.11</td><td>0.3042</td><td>0.2996</td><td>0.3019</td><td>0.245</td></tr><tr><td>33.33</td><td>0.7676</td><td>0.767</td><td>0.7673</td><td>0.7104</td></tr><tr><td>100</td><td>2.0951</td><td>1.9914</td><td>2.0433</td><td>1.9864</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 1.6 | 10.1 | 31.0 | 1.7 | 9.9 | 29.9 |
| Standard Deviation | 0.1 | 0.3 | 0.8 | 0.1 | 0.4 | 1.1 |
| Coefficient of Variation(%) | 6.8 | 2.7 | 2.5 | 6.7 | 3.8 | 3.6 |
The spike recovery was evaluated by spiking 3 levels of human IL-6 into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 84% to 114% with an overall mean recovery of 99%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 2.45-53.36 | 100 | 23.94 |
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.