Human IL-5 ELISA Kit
Sensitivity | 0.15 pg/ml (50 μl) ;0.73 pg/ml (10 μl) |
Standard Curve Range | 1.37-1000 pg/ml |
Standard Curve Gradient | 7 Points/3 Folds |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl/10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | B-cell differentiation factor I, Eosinophil differentiation factor, T-cell replacing factor |
Gene Symbol | IL5 |
Entrez Gene | 3567 (Human) |
SwissProt | P05113 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0324</td><td>0.0303</td><td>0.0314</td><td></td></tr><tr><td>1.37</td><td>0.0392</td><td>0.0346</td><td>0.0369</td><td>0.0056</td></tr><tr><td>4.12</td><td>0.0484</td><td>0.0466</td><td>0.0475</td><td>0.0162</td></tr><tr><td>12.35</td><td>0.0832</td><td>0.0818</td><td>0.0825</td><td>0.0512</td></tr><tr><td>37.04</td><td>0.1765</td><td>0.1662</td><td>0.1714</td><td>0.14</td></tr><tr><td>111.11</td><td>0.4454</td><td>0.4245</td><td>0.435</td><td>0.4036</td></tr><tr><td>333.33</td><td>1.1039</td><td>1.0797</td><td>1.0918</td><td>1.0605</td></tr><tr><td>1000</td><td>2.8645</td><td>2.8548</td><td>2.8597</td><td>2.8283</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 16.3 | 77.9 | 275.8 | 16.4 | 79.9 | 291.5 |
| Standard Deviation | 0.4 | 2.4 | 11.7 | 0.3 | 1.7 | 8.9 |
| Coefficient of Variation(%) | 2.7 | 3.1 | 4.2 | 2.1 | 2.1 | 3.1 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested Twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human IL-5 into health human serum sample. The un-spiked serum was used as blank in these experiments. The recovery ranged from 88% to 115% with an overall mean recovery of 106%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | n.d.-6.72 | 21 | 2.13 |
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
FAQs
New Products
Instruction