Human IL-1R2 ELISA Kit
Sensitivity | 3.24 pg/ml (50 μl);7.35 pg/ml (10 μl) |
Standard Curve Range | 156.25-10000 pg/ml |
Standard Curve Gradient | 7 Points |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl/10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | |
Gene Symbol | IL1R2 |
Entrez Gene | 7850 (Human) |
SwissProt | P27930 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0135</td><td>0.0145</td><td>0.0140</td><td></td></tr><tr><td>156.25</td><td>0.1175</td><td>0.1120</td><td>0.1148</td><td>0.1008</td></tr><tr><td>312.50</td><td>0.2280</td><td>0.2201</td><td>0.2241</td><td>0.2101</td></tr><tr><td>625.00</td><td>0.4698</td><td>0.4395</td><td>0.4547</td><td>0.4407</td></tr><tr><td>1250.00</td><td>0.8763</td><td>0.8636</td><td>0.8700</td><td>0.8560</td></tr><tr><td>2500.00</td><td>1.6210</td><td>1.5660</td><td>1.5935</td><td>1.5795</td></tr><tr><td>5000.00</td><td>2.7980</td><td>2.6830</td><td>2.7405</td><td>2.7265</td></tr><tr><td>10000.00</td><td>4.0427</td><td>4.0936</td><td>4.0682</td><td>4.0542</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 137.9 | 844.8 | 2612.8 | 156.8 | 932.3 | 2846.7 |
| Standard Deviation | 6.4 | 41.2 | 129.3 | 6.7 | 35.4 | 157.6 |
| Coefficient of Variation(%) | 4.7 | 4.9 | 4.9 | 4.2 | 3.8 | 5.5 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess inter-assay precision.
The spike recovery was evaluated by spiking 3 levels of human IL-1R2 into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 75% to 102% with an overall mean recovery of 94%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 3172.90-7423.33 | 100 | 5663.90 |
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
FAQs