Human IFN-β ELISA Kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 0.35 pg/ml (50 μl)
  • Standard Curve Range: 1.37~1000 pg/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AEH0017
US$ Please choose
US$ Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
Product Details
Background
Interferon-beta (IFN-beta) is a cytokine that is produced by fibroblasts and pathogen-exposed dendritic cells, macrophages, and endothelial cells. It signals through the heterodimeric IFN-alpha/beta Receptor. IFN-beta-deficient mice show increased susceptibility to experimental autoimmune encephalomyelitis (EAE), a disease model of human multiple sclerosis (MS). Furthermore, IFN-beta has been shown to suppress the Th17 cell response in both MS and EAE and has commonly been used as a treatment for MS.
Typical data

pg/ml O.D. Average Corrected
0.00 0.0103 0.0106 0.0105  
1.37 0.0160 0.0162 0.0161 0.0057
4.12 0.0305 0.0297 0.0301 0.0197
12.35 0.0641 0.0661 0.0651 0.0547
37.04 0.1846 0.1857 0.1852 0.1747
111.11 0.5026 0.5403 0.5215 0.5110
333.33 1.4850 1.5010 1.4930 1.4826
1000.00 3.5530 3.5690 3.5610 3.5506
Precision
Intra-assay Precision Inter-assay Precision
Sample Number S1 S2 S3 S1 S2 S3
22 22 22 6 6 6
Average(pg/ml) 50.6 281.6 638.2 54.7 293.5 612.2
Standard Deviation 2.7 14.2 28.7 2.3 11.4 35.5
Coefficient of Variation(%) 5.3 5.1 4.5 4.2 3.9 5.8

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human IFN-β into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 75% to 124% with an overall mean recovery of 91%.
Sample Values
Sample Matrix Sample Evaluated Range (pg/ml) Detectable (%) Mean of Detectable (pg/ml)
Serum 30 n.d.-2.08 0.53 1.03

Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of IFN-β in this assay. No medical histories were available for the donors.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.