Human ICAM-3/CD50 ELISA Kit
Sensitivity | 7.09 pg/ml (50 μl);15.79 pg/ml (10 μl); |
Standard Curve Range | 40.96-10000 pg/ml |
Standard Curve Gradient | 7 Points |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl/10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | CDw50; ICAM-R |
Gene Symbol | ICAM3 |
Entrez Gene | 3385 (Human) |
SwissProt | P32942 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0388</td><td>0.0398</td><td>0.0393</td><td></td></tr><tr><td>40.96</td><td>0.0604</td><td>0.0594</td><td>0.0599</td><td>0.0206</td></tr><tr><td>102.40</td><td>0.0958</td><td>0.0936</td><td>0.0947</td><td>0.0554</td></tr><tr><td>256.00</td><td>0.1866</td><td>0.1878</td><td>0.1872</td><td>0.1479</td></tr><tr><td>640.00</td><td>0.3785</td><td>0.3978</td><td>0.3882</td><td>0.3489</td></tr><tr><td>1600.00</td><td>0.8924</td><td>0.8582</td><td>0.8753</td><td>0.8360</td></tr><tr><td>4000.00</td><td>1.9740</td><td>1.9330</td><td>1.9535</td><td>1.9142</td></tr><tr><td>10000.00</td><td>3.5690</td><td>3.5980</td><td>3.5835</td><td>3.5442</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 305.0 | 1111.8 | 3290.7 | 230.0 | 1081.0 | 3271.7 |
| Standard Deviation | 5.7 | 29.7 | 61.8 | 10.0 | 51.1 | 86.0 |
| Coefficient of Variation(%) | 1.9 | 2.7 | 1.9 | 4.4 | 4.7 | 2.6 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human ICAM-3/CD50 into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 98% to 114% with an overall mean recovery of 103%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 93.54-214.96 | 100 | 143.28 |
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.