Human ICAM-1 / CD54 ELISA Kit
Sensitivity | 4.30 pg/ml (50 μl);34.71 pg/ml (10 μl) |
Standard Curve Range | 13.72-10000 pg/ml |
Standard Curve Gradient | 7 Points/3 Folds |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl/10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | Major group rhinovirus receptor |
Gene Symbol | ICAM1 |
Entrez Gene | 3383 (Human) |
SwissProt | P05362 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0128</td><td>0.0150</td><td>0.0139</td><td></td></tr><tr><td>13.72</td><td>0.0217</td><td>0.0241</td><td>0.0229</td><td>0.0090</td></tr><tr><td>41.15</td><td>0.0372</td><td>0.0347</td><td>0.0360</td><td>0.0221</td></tr><tr><td>123.46</td><td>0.0720</td><td>0.0770</td><td>0.0745</td><td>0.0606</td></tr><tr><td>370.37</td><td>0.1703</td><td>0.1777</td><td>0.1740</td><td>0.1601</td></tr><tr><td>1111.11</td><td>0.4687</td><td>0.4474</td><td>0.4581</td><td>0.4442</td></tr><tr><td>3333.33</td><td>1.2790</td><td>1.2190</td><td>1.2490</td><td>1.2351</td></tr><tr><td>10000.00</td><td>2.8090</td><td>2.9300</td><td>2.8695</td><td>2.8556</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 234.1 | 1082.2 | 3166.5 | 213.4 | 1189.1 | 3504.5 |
| Standard Deviation | 19.8 | 43.9 | 70.1 | 17.3 | 46.1 | 211.0 |
| Coefficient of Variation(%) | 8.4 | 4.1 | 2.2 | 8.1 | 3.9 | 6.0 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested Twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human ICAM-1/CD54 into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 97% to 120% with an overall mean recovery of 118%.
| Sample Matrix | Sample Evaluated | Range (ng/ml) | Detectable (%) | Mean of Detectable (ng/ml) |
|---|---|---|---|---|
| Serum | 30 | 46.75-110.67 | 100 | 88.85 |
FAQs
New Products
Instruction