Human Granzyme A/GZMA ELISA Kit
Sensitivity | 0.46 pg/ml (50 μl);1.46 pg/ml (10 μl) |
Standard Curve Range | 2.74-2000 pg/ml |
Standard Curve Gradient | 7 Points/3 Folds |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl/10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | CTL tryptase, Cytotoxic T-lymphocyte proteinase 1, Fragmentin-1, Granzyme-1, Hanukkah factor |
Gene Symbol | GZMA, CTLA3, HFSP |
Entrez Gene | 3001 (Human) |
SwissProt | P12544 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0136</td><td>0.0141</td><td>0.0139</td><td></td></tr><tr><td>2.74</td><td>0.0255</td><td>0.0239</td><td>0.0247</td><td>0.0109</td></tr><tr><td>8.23</td><td>0.0392</td><td>0.0410</td><td>0.0401</td><td>0.0263</td></tr><tr><td>24.69</td><td>0.0845</td><td>0.0824</td><td>0.0835</td><td>0.0696</td></tr><tr><td>74.07</td><td>0.2141</td><td>0.2112</td><td>0.2127</td><td>0.1988</td></tr><tr><td>222.22</td><td>0.6253</td><td>0.5911</td><td>0.6082</td><td>0.5944</td></tr><tr><td>666.67</td><td>1.7910</td><td>1.7440</td><td>1.7675</td><td>1.7537</td></tr><tr><td>2000.00</td><td>3.7830</td><td>3.7900</td><td>3.7865</td><td>3.7727</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 49.6 | 198.7 | 609.8 | 47.5 | 189.0 | 608.9 |
| Standard Deviation | 2.3 | 8.7 | 26.3 | 1.5 | 7.3 | 20.0 |
| Coefficient of Variation(%) | 4.6 | 4.4 | 4.3 | 3.2 | 3.9 | 3.3 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested Twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human Granzyme A into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 81% to 97% with an overall mean recovery of 91%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 62.47-179.55 | 100.0 | 113.58 |
FAQs