Human CD40 ELISA Kit
Sensitivity | 0.34 pg/ml (50 μl) |
Standard Curve Range | 1.37-1000 pg/ml |
Standard Curve Gradient | 7 Points/3 Folds |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | B-cell surface antigen CD40, Bp50, CD40L receptor, CDw40 |
Gene Symbol | CD40, TNFRSF5 |
Entrez Gene | 958 (Human) |
SwissProt | P25942 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0222</td><td>0.0225</td><td>0.0224</td><td></td></tr><tr><td>1.37</td><td>0.0257</td><td>0.0279</td><td>0.0268</td><td>0.0045</td></tr><tr><td>4.12</td><td>0.0406</td><td>0.0415</td><td>0.0411</td><td>0.0187</td></tr><tr><td>12.35</td><td>0.0851</td><td>0.0860</td><td>0.0856</td><td>0.0633</td></tr><tr><td>37.04</td><td>0.2245</td><td>0.2166</td><td>0.2206</td><td>0.1982</td></tr><tr><td>111.11</td><td>0.6181</td><td>0.6157</td><td>0.6169</td><td>0.5946</td></tr><tr><td>333.33</td><td>1.7090</td><td>1.7460</td><td>1.7275</td><td>1.7052</td></tr><tr><td>1000.00</td><td>3.8810</td><td>3.9440</td><td>3.9125</td><td>3.8902</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 23.6 | 121.8 | 373.3 | 23.7 | 115.7 | 342.3 |
| Standard Deviation | 0.5 | 2.8 | 8.2 | 1.0 | 3.0 | 12.5 |
| Coefficient of Variation(%) | 2.1 | 2.3 | 2.2 | 4.1 | 2.6 | 3.7 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human CD40 into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 91% to 121% with an overall mean recovery of 110%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 208.10-361.52 | 100 | 281.37 |
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.