Human CD200R1 ELISA Kit
Sensitivity | 0.1 pg/ml (50 μl);3.09 pg/ml (10 μl) |
Standard Curve Range | 6.86-5000 pg/ml |
Standard Curve Gradient | 7 Points/3 Folds |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl/10 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | CD200 cell surface glycoprotein receptor, Cell surface glycoprotein OX2 receptor 1 |
Gene Symbol | CD200R1, CD200R, CRTR2, MOX2R, OX2R |
Entrez Gene | 131450 (Human) |
SwissProt | Q8TD46 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0140</td><td>0.0153</td><td>0.0147</td><td></td></tr><tr><td>6.86</td><td>0.0195</td><td>0.0207</td><td>0.0201</td><td>0.0055</td></tr><tr><td>20.58</td><td>0.0312</td><td>0.0276</td><td>0.0294</td><td>0.0148</td></tr><tr><td>61.73</td><td>0.0620</td><td>0.0644</td><td>0.0632</td><td>0.0486</td></tr><tr><td>185.19</td><td>0.1621</td><td>0.1541</td><td>0.1581</td><td>0.1435</td></tr><tr><td>555.56</td><td>0.4406</td><td>0.4431</td><td>0.4419</td><td>0.4272</td></tr><tr><td>1666.67</td><td>1.2060</td><td>1.2180</td><td>1.2120</td><td>1.1974</td></tr><tr><td>5000.00</td><td>2.7690</td><td>2.6630</td><td>2.7160</td><td>2.7014</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 1690.5 | 461.7 | 84.6 | 1620.9 | 435.3 | 81.4 |
| Standard Deviation | 121.9 | 24.2 | 4.1 | 109.3 | 27.2 | 5.5 |
| Coefficient of Variation(%) | 7.2 | 5.2 | 4.8 | 6.7 | 6.2 | 6.7 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested Twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human CD200R1 into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 76% to 123% with an overall mean recovery of 102%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 747.49-1789.41 | 100 | 1147.83 |
FAQs
New Products
Instruction