Human Apolipoprotein A1/APOA1 ELISA Kit
Sensitivity | 91.61 pg/ml (50 μl) |
Standard Curve Range | 468.75-30000 pg/ml |
Standard Curve Gradient | 7 Points |
Number of Incubations | 2 |
Detectable Sample | Liquid phase sample of soluble substances. For example: serum, plasma, cell culture supernatant, tissue grinding liquid, etc. |
Sample Volume | 50 μl |
Type | Fully Ready-to-Use |
Operation Duration | 120min |
Alternative Names | Apolipoprotein A1 |
Gene Symbol | APOA1 |
Entrez Gene | 335 (Human) |
SwissProt | P02647 (Human) |

<table><tr><td>pg/ml</td><td>O.D.1</td><td>O.D.2</td><td>Average</td><td>Corrected</td></tr><tr><td>0.00</td><td>0.0091</td><td>0.0093</td><td>0.0092</td><td></td></tr><tr><td>468.75</td><td>0.0140</td><td>0.0140</td><td>0.0140</td><td>0.0048</td></tr><tr><td>937.50</td><td>0.0294</td><td>0.0286</td><td>0.0290</td><td>0.0198</td></tr><tr><td>1875.00</td><td>0.0781</td><td>0.0818</td><td>0.0800</td><td>0.0708</td></tr><tr><td>3750.00</td><td>0.2654</td><td>0.2686</td><td>0.2670</td><td>0.2578</td></tr><tr><td>7500.00</td><td>0.8807</td><td>0.8709</td><td>0.8758</td><td>0.8666</td></tr><tr><td>15000.00</td><td>2.1960</td><td>2.2280</td><td>2.2120</td><td>2.2028</td></tr><tr><td>30000.00</td><td>4.4280</td><td>4.2416</td><td>4.3348</td><td>4.3256</td></tr></table>
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 411.0 | 2699.3 | 8436.9 | 609.6 | 2786.8 | 8924.8 |
| Standard Deviation | 24.2 | 110.0 | 280.3 | 21.2 | 15.1 | 427.6 |
| Coefficient of Variation(%) | 5.9 | 4.1 | 3.3 | 3.5 | 0.5 | 4.8 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The spike recovery was evaluated by spiking 3 levels of human Apolipoprotein A1/APOA1 into health human serum sample. The un-spiked serum was used as blank in this experiment. The recovery ranged from 93% to 120% with an overall mean recovery of 107%.
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 683.85-2777.78 | 100 | 1035.81 |
n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
FAQs
New Products
Instruction